Mobile Magnetic Resonance Imaging System (MRI) and MRI Technologist Lease

Status
Closed, the response deadline passed on Mar 28, 2025
Agency
United States Department of Veterans Affairs
Office
ARLINGTON, USA
Notice type
Sources Sought
Solicitation number
36C257-25-AP-1637
NAICS
532490 - Other Commercial and Industrial Machinery and Equipment Rental and Leasing
PSC
W065 - LEASE OR RENTAL OF EQUIPMENT- MEDICAL, DENTAL, AND VETERINARY EQUIPMENT AND SUPPLIES
Category
Medical Equipment and Accessories and Supplies (10.2)
Posted
2025-03-24
Deadline
2025-03-28
Place of performance
Big Spring, Texas 79720, UNITED STATES

Description

  1. Scope: The intent of this Request for Proposal is to enter into a contract to provide Mobile MRI and Technologist Services to include all the necessary personnel, supervision, equipment, transportation, materials, supplies and training to perform the services as specified herein. Service shall be performed at any West Texas VA Healthcare Facility, including both owned and/or leased facilities, main and/or community-based outpatient clinics where indicated.

  1. Summary of Requirement: The West Texas Healthcare System is seeking a 12-month Base, plus 4 (four) 12-month Option Periods, contract for a mobile MRI and associated service agreement and Technologist staffing. The mobile MRI will be placed on an as-needed basis for delivery or move within 10 days, to remain in a single location no less than 6 months.

  1. Delivery and Set-up of Contractor Equipment:  The Contractor shall be responsible for proper installation/set-up of the Contractor's furnished equipment including the mobile coach. Installation/set-up shall be accomplished in accordance with industry standards, all OSHA regulations and applicable manufacturer's recommendations.

  

    1. The Contractor shall provide a mobile 1.5T MRI coach and MRI Technologists Services at station 519 (West Texas Healthcare System), in accordance with the SOW (Statement of Work).

    1. Schedule:  Services shall be required five (5) days per week, except on federal holidays.

    1. Hours of Service:  Contractor's mobile unit and/or personnel shall be on site to perform exams eight (8) hours per day, Monday - Friday, between the hours of 7:30 am and 4:30 pm. The first MRI will begin at 7:30 am with the last patient being seen such that the exam is completed no later than 4:30pm. 

    1. Pro-rated Monthly Rental Fee: The Contractor shall prorate the monthly rental fee when MRI coach is not functioning in accordance with prescribed guidelines.  The monthly fee shall be prorated based on the number of hours the coach is out of service during regularly scheduled clinic hours M-F, 7:30am – 4:30pm.  The Contractor shall include the hourly rate in the space provided below:

  1. Contractor's Responsibilities:  The Contractor shall be responsible for the following daily functions:

    1. Contractor shall be responsible for the transportation of the Mobile MRI coach to and from WTVAHCS. Cost of transportation shall be included in the monthly rate. Big Spring facility is gated and locked after-hours. The facility is open from 6:30 am through 6 pm Monday through Friday for removal of trailer for other assignments.

    1. Delivery and Set-up of Contractor Equipment: The Contractor shall be responsible for proper installation/set-up of the Contractor's furnished equipment including the mobile trailer. Installation/setup shall be accomplished in accordance with industry standards, all OSHA regulations and applicable manufacturer's recommendations.

    1. Transmit completed electronic data after each exam.  Ensure that all of the day’s exam data is completely transferred to the West Texas VA Healthcare System computer systems.

    1. All patient information required for Contractor's records shall be secured in accordance with HIPAA guidelines.

    1. All connections from the trailer to West Texas VA Healthcare System site/property are properly disconnected from trailer when moved from the site.

    1. Any temporary site preparations are removed upon completion of services.

    1. Contractor shall ensure mobile MRI is maintained in optimum working condition to ensure the highest level of patient safety at all times. Fire extinguisher to be supplied by the vendor. Vendor shall provide MRI compatible wheelchair, stretcher and O2 tank.

    1. Maintenance documents shall be made available to the West Texas VA Healthcare System upon request.

    1. Contractor shall ensure each patient arriving for mobile MRI exam has a signed MRI order for services.

    1. MRI tech must complete the final MRI Screening Assessment which is incorporated in the patient’s electronic medical record.

    1. Contractor shall use appropriate screening devices to ensure patient has no ferro-magnetic materials on or in receive his/her body.

    1. Contractor shall be responsible for adhering to all Federal regulations regarding patient care, patient safety, medical practice and operation of mobile MRI.

    1. Contractor shall provide a copy of a current, comprehensive MRI Safety Program policy to which each technologist is expected to comply. MRI Technologist is required to complete West Texas VA Healthcare System or analogous MRI Safety Training yearly.

    1. Contractor shall be responsible for patient while on the mobile unit and during the entire scanning procedure up to the patient's release.

    1. Environment of Care Rounds shall be conducted quarterly by the local site COR and West Texas VA Healthcare System Administrator. All identified deficiencies shall be corrected by the Contractor prior to the next scheduled day of service.

    1. Contractor shall be responsible for providing 2 qualified MRI Technologists to perform services defined herein. The MRI Technologists shall be required to provide scanning services as well as training on the full operation of the Mobile MRI unit. Training shall be provided to the assigned West Texas VA Healthcare System technical personnel at no additional cost to the Government. If the MRI Technologist is unable to come in to work as per the schedule, an alternate MRI Technologist is available to cover the shift to avoid having to reschedule patients. All MRI Technologists must complete the credentialing requirements established by West Texas VA Healthcare System.

    1. MRI Technologist shall attend the quarterly MRI Safety Committee meeting.

    1. Comply with VHA Directive 1105.05 Magnetic Resonance (MR) Safety, WTVAHCS MRI Policy and Procedures and MRI scanning protocols.

5. Delivery Locations:

           

            See below for location of West Texas VA Healthcare System sites of care. The delivery locations include but are not limited to these locations. If new VA locations within the geographic boundaries of West Texas VA Healthcare System (such as a new clinic in Abilene) are added during the duration of the contract, those will presume to be included as potential deployment locations for a mobile MRI as well, provided that they are affiliated with station 519.

George H. O'Brien, Jr., Department of Veterans Affairs Medical Center

300 Veterans Boulevard

Big Spring, TX 79720-5566

Abilene VA Clinic

3850 Ridgemont Drive

Abilene, TX 79606-2728

Colonel Charles and JoAnne Powell VA Clinic

4240 Southwest Boulevard

San Angelo, TX 76904-5634

Fort Stockton VA Clinic

1205 North Sycamore Street

Fort Stockton, TX 79735-4122

Hobbs VA Clinic

1301 North Turner Street

Hobbs, NM 88240-8446

Wilson and Young Medal of Honor VA Clinic

8050 East Highway 191

Odessa, TX 79765-8615

8. Magnetic Resonance Imaging (MRI) Mobile Device Salient Characteristics:

Contractor shall make whatever adaptations are necessary so that the Contractor's equipment is compatible with the West Texas Healthcare System data network as specified in Paragraph 16, Telecommunications.

Salient characteristics

Details

Magnet strength 

1.5 Tesla 

Minimum Bore width (cm) 

70cm 

Minimum bore depth (cm) 

130cm 

Minimum slew rate (T/m/s) 

125 

Minimum number of channels 

48 

Maximum patient table load capacity for vertical and horizontal table movement (lbs) 

550 

Maximum system weigh (lbs) 

10,000 

Helium consumption 

Zero Helium boil-off technology 

Min patient comfort requirement 

Short and wide bore for greater patient comfort and easier access for various patient sixes 

Gradient system 

XJ Gradients with high performance capable of handling advanced imaging techniques 

RF Technology 

TIM 4G 9Toal Imaging Matrix) technology, allowing flexible coil combinations and faster imaging 

Coil Technology 

Up to 204 coil elements, allowing for whole body imaging without patient repositioning 

Imaging speed 

Up to 48 channels supported with Tim 4G, enabling faster scans and higher throughput 

Image quality 

High spatial and temporal resolution, with advanced features, post processing and workflow optimization tools 

Patient handling 

Comfortable positioning for long exams, with respiratory and cardiac gating options 

Application range 

Neurology, musculoskeletal, cardiovascular, Ortho, Oncology, MRA, Elastography, Spectroscopy, Phase contrast MRI and flow analysis, parametric prostate imaging, high intensity focused ultrasound with the flexibility to expand due to different facility provider requirements 

Advanced Imaging Techniques 

Diffusion weighted imaging (DWI), functional MRI (fMRI), cardiac imaging, MR spectroscopy, etc. 

Environment  

Energy efficiency and reduced operating costs 

Installation space 

Mobile trailer installation capable; Semi-trailer truck equipped with stabilizing shocks to prevent disruption of the machine, compact design with low space requirements ideal for a mobile trailer 

Operator workflow 

Intuitive user interface, customizable protocols, and easy data management 

Connectivity 

DICOM compliant integrated with hospital information system (HIS) and Picture Archiving (PACS) 

Upgradeable  

Must be upgradeable to newer software and hardware innovations 

Safety features 

Advanced patient monitoring, acoustic noise reduction, and patient emergency systems. Safety system for magnet quenching and emergency shutdown, to include the following: 

Table requirement  

UPS requirement 

Maintains full functionality of the system for a minimum of 30 minutes during a power outage 

Coil requirement 

Imaging of the Torso, knee, shoulder, head/spine/array, neuro vascular, run-off, foot/ankle, wrist, elbow, cardiac, breast, prostate imaging, functional MRI head, carotid, endorectal, periphery vascular, extremity, flexible, transmit/receive, adjustable neurovascular head, whole body imaging 

Security/Connectivity requirements  

Chiller requirement  

Chiller meeting or exceeding MRI needs 

Cryogen requirement 

Cryogens (Vendor expected to provide and install all cryogens.) 

Phantom requirements 

All phantoms required for proper calibration and performance verification (Vendors are expected to provide phantoms for all system requirements and options identified in this document that require phantoms.) 

Patient comfort/entertainment package 

Integrated music system, to include the following: 

Shielding requirement 

Active Shielding 

Workstation requirement 

Additional technology specifications 

  1. The Mobile MRI unit shall be ready for use regardless of outside environmental conditions.
  2. The Mobile MRI unit shall maintain a temperature to assure proper operation of the scanner and provide for patient comfort.
  3. The Mobile MRI unit shall include a private dressing area with secured cabinets for patient use.
  4. Contractor shall provide an MRI-compatible Automatic External Defibrillator (AED).  The system must contain the minimum characteristics:
    1. AED Unit with LCS Screen showing voice prompt messages, device advisory messages, elapsed time, shock count and chest compression graph, Operator's Guide and Carrying Case;
    2. Two (2) Sets of CPR-D Pads, one piece defibrillation and CPR system with compression, depth and rate sensors, supplies with gloves, barrier mask, scissors, razor, wet wipes and dry wipes.

  

Trailer space requirement 

(based on max system dimensions) 

Location 

Requirement 

Width (cm) 

Depth (cm) 

Height (cm) 

Weight (cm) 

Examination Room 

Magnet and Helium 

210 

150 

220 

3200 

Magnet, gradient coils body coil, table and covers 

250 

450 

225 

5000 

Min room height clearance 

245 

Control Room 

Workplace table and monitor 

130 

90 

120 

Equipment room 

Cabinets with system controls, RF system, gradient power system image processor 

170 

70 

200 

1500 

9. MRI Technologist Staffing Salient Characteristics:

a.    The Contractor's Technologists shall possess a current American Registry of Radiologic Technologists (ARRT) license, certified in MRI through ARRT and Basic Life Support (BLS) certified.  A copy of all applicable operators’ license/certifications shall be provided with the proposal along with a copy of the updated license immediately upon renewal.  The Contractor shall ensure that a copy of the technologist’s current license is provided to the local COR for inclusion in the West Texas VA Healthcare System 6-part folder at all times.

b.   The Contractor shall be responsible to ensure that Contractor employees providing work on this contract are fully trained and completely competent to perform the required work.  Evidence of the Contractor’s Technologist’s competency review shall be provided with the proposal.  Competency checklists must contain evidence of supervisory review at least annually and must contain the written signature of the supervisory official performing the review.  A current copy of the employee’s competency checklist shall be maintained in West Texas VA Healthcare System 6-part folder at all times.  Competency checklists shall