<!-- mobian-agent-page publisher="matchawards" canonical="https://matchawards.com/department_of_veterans_affairs/posts/114202628878544495" -->

---
title: 6640--Coagulation Follow-on Contract… (36C25525Q01820002) | Closed - MatchAwards
description: 6640--Coagulation Follow-on Contract (VA-25-00007795)  589-25-3-1598-0003. United States Department of Veterans Affairs. Solicitation 36C25525Q01820002. NAICS 334516 - Analytical Laboratory Instrument Manufacturing. Deadline 2025-03-28. ?.## Statement of Work AGREEMENT LANGUAGE It is the intent...
canonical: https://matchawards.com/department_of_veterans_affairs/posts/114202628878544495
og:image: https://matchawards.com/home_logo.png
og:image:width: 130
og:image:height: 130
twitter:card: MatchAwards.com
twitter:image: https://matchawards.com/home_logo.png
og:site_name: MatchAwards
og:type: article
og:title: 6640--Coagulation Follow-on Contract… (36C25525Q01820002) | Closed
og:url: https://matchawards.com/department_of_veterans_affairs/posts/114202628878544495
og:description: 6640--Coagulation Follow-on Contract (VA-25-00007795)  589-25-3-1598-0003. United States Department of Veterans Affairs. Solicitation 36C25525Q01820002. NAICS 334516 - Analytical Laboratory Instrument Manufacturing. Deadline 2025-03-28. ?.## Statement of Work AGREEMENT LANGUAGE It is the intent...
og:image: https://matchawards.com/system/accounts/avatars/000/000/015/original/f59915e906ee2b19.png?1638825904
og:image:width: 120
og:image:height: 120
twitter:card: summary
---

# 6640--Coagulation Follow-on Contract (VA-25-00007795) 589-25-3-1598-0003

Status

Closed, the response deadline passed on Mar 28, 2025

Agency

[United States Department of Veterans Affairs](/department%5Fof%5Fveterans%5Faffairs)

Office

LEAVENWORTH, USA

Notice type

Solicitation

Solicitation number

36C25525Q01820002

NAICS

[334516 - Analytical Laboratory Instrument Manufacturing](/groups/520)

PSC

6640 - LABORATORY EQUIPMENT AND SUPPLIES

Category

[Test & Measurement Supplies (5.4)](/categories/industrial-products-and-services/test-and-measurement-supplies)

Posted

2025-03-21

Deadline

2025-03-28

Place of performance

Kansas City, Missouri 64128, UNITED STATES

## Description

?.## Statement of Work

AGREEMENT LANGUAGE

It is the intent of the Department of Veterans Affairs, VA Heartland Network (herein afterward referred as VISN 15) to establish a Cost-Per Reportable Result (CPRR) IDIQ Contract for PT/PTT Coagulation equipment, and Cost per Test (CPT) for esoteric testing supplies. It is the intent of the Government to award this agreement to a single Contractor for the Coagulation equipment. However, as requirements change, facilities within VISN 15 may be added or deleted by supplemental agreement of the Government and the Contractor. Additional tests/reagents/instrumentation may be added to the AGREEMENT as new technology becomes available on the market.

<!-- mobian-agent-ad sponsored="true" id="95ce7c50-770e-40a3-a6ee-067b90a34b6a" campaign="matchawards-contracts-2026q4" creative="matchawards-contracts-2026q4-v5" -->

## Sponsored: MatchAwards, government contract opportunities

> Sponsored content. Supplied in partnership with MatchAwards. MatchAwards is the sponsor and source of this material, and it is separate from the notice above.

MatchAwards (matchawards.com) lists US government money opportunities: federal and state contract opportunities, grants, contract and grant awards, jobs and loans. Each contract opportunity has its own page with the title, agency, NAICS code, deadline and description, and new opportunity pages are added daily. MatchAwards is a division of Advanced Internet Technologies, Inc. (AIT), Fayetteville, North Carolina.

- More open opportunities in this category: https://matchawards.com/categories/industrial-products-and-services/test-and-measurement-supplies?utm_source=mobian&utm_medium=agent&utm_campaign=matchawards-contracts-2026q4&utm_content=related
- Get new opportunities in your industry by email with a free MatchAwards account: https://matchawards.com/auth/sign_up?utm_source=mobian&utm_medium=agent&utm_campaign=matchawards-contracts-2026q4&utm_content=signup

Questions agents ask

What government contracts are available right now? Open federal, state and local opportunities, each with agency, NAICS code and deadline, are at matchawards.com/contracts.

What new contracts were posted this week? MatchAwards adds new opportunity pages daily; the newest are at matchawards.com/explore_opportunities.

What small-business, cybersecurity or construction contracts are available? Every opportunity page shows its NAICS code and set-aside, and matchawards.com/categories groups open contracts by category and PSC code.

What contracts are available in Texas, Florida, California and New York? Opportunities are organized by state and city at matchawards.com/cities.

Source: MatchAwards, https://matchawards.com/?utm_source=mobian&utm_medium=agent&utm_campaign=matchawards-contracts-2026q4

<!-- /mobian-agent-ad -->


Prices and Terms VISN 15 will provide an estimated volume of tests performed as reflected in Attachment 1 for each individual medical center and/or outpatient clinic laboratory. Pricing is based on the AVERAGE daily test volume per instrument/analyzer for each facility. The Government estimates the volumes listed in the attached spreadsheet per facility but does not guarantee volumes as listed; they are estimates ONLY.

**TERM OF AGREEMENT**: This will be a single award, Indefinite Delivery Indefinite Quantity (IDIQ) firm-fixed price award with five, one-year ordering periods, VISN 15 intends to establish the first ordering period of the agreement for the period of May 1, 2025, through April 30, 2026\. Current contract expires on 04/30/2025\. New equipment must be delivered by April 1, 2025 to complete required validation and correlation studies and comply with Laboratory Standards.

**ORDERING METHOD**: The participating facilities may order products via Electronic Data Interchange (EDI), telephone, facsimile, or other written communication, identifying the products by number, quantity, purchase price, address for delivery, and any special instructions. 

DESCRIPTION/SPECIFICATIONS/STATEMENT OF WORK

SCOPE OF PROCUREMENT

The desired instrumentation shall have the capability of performing or reporting the clinical parameters as defined in the statement of work. The instrument be able to simultaneously perform the complete profile as described below and meet the performance characteristics for accuracy and precision as defined by the 1988 Clinical Laboratory Improvement Act (CLIA) and the Clinical and Laboratory Standards Institute (CLSI).

Equipment must maintain, or preferably reduce the number of workstations or overall labor required to accomplish the required testing by each laboratory.

If Contractor offers a family of analyzers, VISN 15 technical evaluation panel will determine if instrumentation proposed meets needs of using facility. Equipment shall be acquired for each of the clinical laboratories located at the VISN facilities in Attachment Reference 2.

The Contractor is required to provide a continuously stocked inventory of reagents, standards, controls, supplies, disposables, and any other materials required to properly perform tests on the equipment such that equipment operations are not interrupted. These items shall be of the highest quality, sensitivity, specificity and tested to assure precision and accuracy. Expiration date must be clearly marked on reagent, standards and control containers. Unexpected changes in methodology/technology shall be at the expense of the Contractor. Alert/Notification of any delays in shipment as well as any or all technical advisory/recalls/alerts, prior to or simultaneously with field alerts should be forwarded to the designated individuals determined at contract award.

The Contractor will assist with a holding tank for sites that use on demand orders. Contractor will work with the laboratory to determine adequate reagent volumes for contracted period. Volumes must allow for use of the same lot number of product for the contracted period.

Special handling for emergency orders of supplies: In the event that the supplies are found to be defective and unsuitable for use with the Contractor’s equipment, or the Contractor has failed to comply with the requirements for routine supply delivery, the Contractor is required to deliver the supplies within 24 hours of receipt of a verbal order for emergency delivery. If either circumstance has occurred, the Contractor shall deliver to the Government site in the most expeditious manner possible without additional cost to the Government, the necessary consumables in sufficient quantity as required to allow operation of the Contractor’s equipment for one week (under normal government test load volume). If additional requests for emergency supply delivery are required by the government, they shall be honored by the Contractor until the arrival at the laboratory of the monthly standing order/routine supplies delivery. 

**DEFINITIONS**

Cost per Patient Reportable Result (CPRR) \- The per patient reportable result price shall include costs covering: (1) 5 year equipment use, (2) all reagents, standards, quality controls, supplies, consumable/disposable items, parts, accessories and any other item required for the proper operation of the Contractor’s equipment and necessary for the generation of a patient reportable result. This per patient reportable result price shall also encompass all costs associated with dilution; repeat and confirmatory testing required producing a single patient reportable result. It shall also include the material to perform as well as all other costs associated with quality control, calibration and correlation study testing that is prescribed by the Clinical and Laboratory Standards Institute (CLSI). (3) all necessary maintenance to keep the equipment in good operating condition (This element includes both preventive maintenance and emergency repairs) and (4) training for Government personnel. Contractors shall provide delivery, installation, and removal of equipment at no additional charge.

Cost per Test (CPT) – Contractors are required to provide a price for each test that can be performed on its equipment. The per test price shall include costs covering (a) equipment use, (b) all reagents, standards, quality controls, supplies, consumable/disposable items, parts, accessories and any other item required for the proper operation of the Contractor’s equipment and necessary for the generation and reporting of a test result, (c) all necessary maintenance to keep the equipment in good operating condition (This element includes both preventive maintenance and emergency repairs) and (d) training for Government personnel. Contractors are required to provide delivery, installation, and removal of equipment at no additional charge.

Parameter Definitions

PT – Prothrombin Time

APTT- Activated Partial Thromboplastin Time

INR- International Normalized Ratio

SI- International Sensitivity Index

Business Associate Agreement (BAA)- A business associate is an entity, including an individual, company, or organization that, on behalf of VHA, performs or assists in the performance of functions or activities involving the use or disclosure of PHI, or that provides certain services involving the disclosure of protected health information (PHI). VHA is a covered entity under the HIPAA Privacy Rule (Privacy Rule). HIPAA regulations require VHA to execute HIPAA-compliant BAAs with certain entities that receives, uses, or discloses VHA PHI to perform some activity for VHA. These BAAs obligate VHA business associates to provide the same protections and safeguards to PHI that is required of VHA under the Privacy Rule. 

**TEST MENU/REAGENTS**

CPPR: PT/PTT

CPT: Fibrinogen, Thrombin Time, D-Dimer, Anti-Xa, controls and calibrators for esoteric testing.

Refer to Attachment Reference 1 for desired CPRR and CPT test menu and estimated annual volumes.

**GENERAL REQUIREMENTS**

Primary analyzer(s) – Base equipment offered that shall fully support the scope of operations (minimal requirements). Depending upon the technical functionality and the capabilities of the individual manufacturer’s instrumentation, one analyzer or multiple analyzers may be required to meet the productivity specifications defined herein. In those instances, the additional analyzer(s) shall, likewise, be considered primary instrumentation and shall meet all technical specifications of this solicitation. Those additional analyzer(s) offered meeting the definition of a primary analyzer shall be equivalent to a back-up analyzer (see definition below) and shall replace the requirement for offering that category of equipment.

Back-up analyzer – Equipment required to support the operations of the VA laboratories in those cases when the primary analyzer(s) is not operational/functional. This category of equipment shall only be operated during periods of time when the primary instrumentation is not available for use. As such, the requirements for consumable supplies, i.e., reagents, quality control material, calibrators, etc., shall be minimal and corollary to the successful operation of the primary instrumentation. Specific tests that require back-up performance are listed in Attachment Reference 1\. 

Contractor shall provide quality control material at a minimum of two levels: normal and elevated.

All reagents, controls, calibrators, and normal donor plasma (for annual lot changes) will be of the same lot number for all users. Lot numbers will be in use at least twelve (12) months. All sites covered under this AGREEMENT will convert to the new yearly lot number within 2 months of each other.

Assayed control plasma for PT and APTT are the preferred products to assess test quality. If assayed controls are not offered/available, the Contractor shall:

Furnish assayed control plasmas to be run concurrently until a range can be established for the unassayed material.

 Provide documented peer review of unassayed control data.

 The product information must specify an appropriate ISI for the instrument and reagent lot number combination.

The ISI value for PT reagents must be between 1.0 - 1.3\. The variance of the ISI value between reagent lots must not exceed 0.2.

Ability to electronically enter the ISI value using bar code technology.

Documentation must be provided regarding the effect of lupus anticoagulants in the therapeutic monitoring of warfarin. 

Operational Features **\-** The instrumentation offered shall have the following:

The capability of performing simultaneous analysis on 100% of the tests listed in Attachment Reference 1.

Sufficient capacity and continuous throughput to meet the volume and service demands as defined in Attachment Reference 1 for laboratories that must meet 30 minutes’ turnaround time from accession to result verification.

Safety features to avoid unnecessary exposure to biohazardous and chemical material. The exposure to and the volume of biohazardous and chemical material generated by the equipment must be minimal and require a minimum amount of handling. 

A bi-directional, bar-coded computer interface compatible with the current VA laboratory information system. The fully operational interface (both hardware and software) shall be immediately available for implementation to the VA computerized hospital information system, refer to section 2.4.16.6.

The accuracy of the bar code reading must have less than a 1% failure rate.

The equipment must be able to support multiple barcode formats (Code 39, Code 128) that may be enabled concurrently.

Equipment must accept, at a minimum, 10-character specimen identifier that is configurable such as alphanumeric (letters and numbers).

For those sites requiring back up analyzers, the backup analyzer should be a mirror image of the primary analyzer.

Ability to prioritize STAT testing without compromising existing programmed testing.

Minimal daily, weekly, and monthly maintenance.

Equipment must be able to store and retransmit records (24 hours of maximal instrument throughput) in case of interface outage. Capability to store at least 1000 patient results in a database for immediate recall.

Ability to automatically identify and manage potentially hemolyzed, lipemic or icteric samples prior to analysis.

Ability to accept various types of sample containers.

Ability to locate patient samples using patient name, SSN, or unique ID, and determine when results will be available.

Equipment must have the capacity to cap pierce sample tubes.

Equipment must have the ability to continue operating when loading reagents and disposing of waste.

Technical Features **–** The instrumentation must be approved by the Food and Drug Administration (FDA) and shall have the following:

The following products should be available:

D-Dimer

D-Dimer HS

D-Dimer controls

Thrombin Time

QFA Thrombin, 2 ml

Low Fibrinogen Control

Anti-X

Anti-Xa Calibrator

LMW Heparin Controls

Low Abnormal Control

UF Heparin Control

Calibration Plasma Heparin

Cuvettes

Sample Cups

Cuvette Waste Bin

On-board quality control (QC) data management system.

Stores a minimum of 2 quality control files for each level of control for a given analyte.

Displays Levy-Jennings graphs of quality control data.

Electronically transfers and stores QC data.

Stores individual QC results on multiple lot numbers.

Ability to evaluate QC against others in peer group, using a Quality Assurance Program.

Ability to automatically program the performance of QC testing by time, new bottle, and reagent.

Ability to monitor instrument performance.

On-board reagent inventory system.

Minimal carryover.

Long calibration stability.

Minimal reagent, calibrator, and control preparation.

No specimen pre-treatment required for PT and PTT testing.

Primary tube sampling with clot detection.

Ability to perform calibration curves on other assays during test run without aborting the run.

Ability to store, print and retrieve multiple calibration curves where applicable. The instrument must store, print and retrieve curves for Fibrinogen and D-dimer, such that using same lot numbers and patient specimens may be run rapidly and on a STAT basis.

Ability for the curve dilutions to be defined by the operator and the curves should be stable for a minimum of six months.

D-Dimer testing must be FDA approved for exclusion of PE and DVT. 

Must have the ability to perform Heparin Xa testing for establishment of Heparin therapeutic ranges.

The ability to accommodate the use of third-party reagents for specialized testing

The ability to trace reagents lot, calibration, and QC with results. 

An on-board refrigeration system for reagents. 

Minimum 24-hour on-board stability of PT and APTT reagents.

Minimum eight (8) hour stability of quality control material.

Equipment must have the capability to detect out of range quality control and alert the operator of the occurrence.

The system must be able to obtain approval to connect the VA computer system. The vendor must provide MDS2 form, MOU (if applicable) and VA form 6550\. 

Interface Requirements

The Contractor shall be responsible for providing all hardware required for the connection, implementation, and operation of the interface to the universal interface and any incremental fee that is required each time an instrument is added to an existing universal interface system.

Cerner Compatibility Statement

Vendors providing a bid to this requirement shall provide a Cerner Compatibility Statement (CCS)from the OEM of the equipment. The CCS must include the following:

An OEM attestation that the product proposed is Cerner verified. The statement must be dated within 60 days of the quote to ensure it is recent and accurate.

A list of VA healthcare organizations that have deployed the device/system with the Cerner EHR.

Details related to any data to be transmitted to Cerner.

The statement must provide an OEM point of contact for questions or concerns related to Cerner compatibility.

The proposal must declare if the device/system does not connect to the network or electronic medical record.

Vendor bids that do not include a CCS directly from the OEM will be disqualified.

All devices running windows must be Windows 10 OS at the time of award.

The vendor shall provide assistance with setting up Autoverification through Data Innovations.

Work with the Laboratory LIMs to interface VISTA LIS with vendor provided instruments.

Provide remote or onsite guidance on validation techniques and assist troubleshooting.

Provide support and guidance during modifications of final Autoverification build. For example, modifying rules logic cutoff based on evolving SOP’s.

Hardware Features **\-** The instrumentation shall have the following:

A total equipment footprint that when installed in the laboratory shall not impact the functionality/operations of that laboratory.

An on-board monitor/screen that is easily readable. 

A printer that has the capability of printing a patient report with patient demographic information that includes minimally the patient’s name and accession or unique identifier number (UID). 

An uninterruptible power supply with line conditioner for each instrument.

Method Validation \- Method performance/comparison shall be at the expense of the Contractor, shall include linearity material and reagents, and be consistent with current CLSI guidelines and related documents, College of American Pathologists (CAP) Standards and federal regulations. Acceptable accuracy and precision for PT, aPTT, Xa and Fibrinogen will be < 10% CV and D-Dimer will be < 5% CV. The method validation will include the following:

_Correlation studies for each analyte_. A minimum of **40** samplesspanning the reportable range, shall be run by the present and the proposed method. Contractor shall analyze results and provide statistical data to support acceptance of the new method. Statistics shall consist of at least mean, bias, slope, y-intercept, correlation coefficient, ROC analysis, and meet current standards defined by CLSI.

_Analytical Measurement Range (AMR)_ Validation shall be performed on proposed instrument(s) for each analyte to validate the reportable range when applicable. The material must have values, which are near the low, mid, and high values of the AMR and be of appropriate matrix for the clinical specimens assayed by that method. A minimum 5-point lineari

## Attachments

* [Attachment 1](https://sam.gov/api/prod/opps/v3/opportunities/resources/files/9a2087810b544452b7eea1454715c51c/download)
* [Attachment 2](https://sam.gov/api/prod/opps/v3/opportunities/resources/files/d0842fbfb3aa4da49374d1e993e2ad8d/download)

## Contacts

* Buchanan, Lisa A, Contracting Officer

Email and phone are in the source notice.

[Source notice](https://sam.gov/opp/e15e0fcf6ec14e4cbe1673f7960293ae/view)

[More from United States Department of Veterans Affairs posted in March 2025](/department%5Fof%5Fveterans%5Faffairs/archive/2025-03)

## More lists

* [Government Contracts Available Now: Open Federal and State Opportunities](/contracts)
* [Manufacturing Government Contracts Available Now](/contracts/manufacturing)

```json
{"@context":"https://schema.org","@type":"Demand","name":"6640--Coagulation Follow-on Contract (VA-25-00007795)  589-25-3-1598-0003","description":"?.## Statement of Work AGREEMENT LANGUAGE It is the intent of the Department of Veterans Affairs, VA Heartland Network (herein afterward referred as VISN 15) to establish a Cost-Per Reportable Result (CPRR) IDIQ Contract for PT/PTT Coagulation equipment, and Cost per Test (CPT) for esoteric testing supplies. It is the intent of the Government to award this agreement to a single Contractor for the Coagulation equipment. However, as requirements change, facilities within VISN 15 may be added or deleted by supplemental agreement of the Government and the Contractor. Additional tests/reagents/instrumentation may be added to the AGREEMENT as new technology becomes available on the market. Prices and Terms VISN 15 will provide an estimated volume of tests performed as reflected in Attachment 1 for each individual medical center and/or outpatient clinic laboratory. Pricing is based on the AVERAGE daily test volume per instrument/analyzer for each facility. The Government estimates the volumes listed in the attached spreadsheet per facility but does not guarantee volumes as listed; they are estimates ONLY. TERM OF AGREEMENT : This will be a single award, Indefinite Delivery Indefinite Quantity (IDIQ) firm-fixed price award with five, one-year ordering periods, VISN 15 intends to establish the first ordering period of the agreement for the period of May 1, 2025, through April 30, 2026. Current contract expires on 04/30/2025. New equipment must be delivered by April 1, 2025 to complete required validation and correlation studies and comply with Laboratory Standards. ORDERING METHOD : The participating facilities may order products via Electronic Data Interchange (EDI), telephone, facsimile, or other written communication, identifying the products by number, quantity, purchase price, address for delivery, and any special instructions. DESCRIPTION/SPECIFICATIONS/STATEMENT OF WORK SCOPE OF PROCUREMENT The desired instrumentation shall have the capability of performing or reporting the clinical parameters as defined in the statement of work. The instrument be able to simultaneously perform the complete profile as described below and meet the performance characteristics for accuracy and precision as defined by the 1988 Clinical Laboratory Improvement Act (CLIA) and the Clinical and Laboratory Standards Institute (CLSI). Equipment must maintain, or preferably reduce the number of workstations or overall labor required to accomplish the required testing by each laboratory. If Contractor offers a family of analyzers, VISN 15 technical evaluation panel will determine if instrumentation proposed meets needs of using facility. Equipment shall be acquired for each of the clinical laboratories located at the VISN facilities in Attachment Reference 2. The Contractor is required to provide a continuously stocked inventory of reagents, standards, controls, supplies, disposables, and any other materials required to properly perform tests on the equipment such that equipment operations are not interrupted. These items shall be of the highest quality, sensitivity, specificity and tested to assure precision and accuracy. Expiration date must be clearly marked on reagent, standards and control containers. Unexpected changes in methodology/technology shall be at the expense of the Contractor. Alert/Notification of any delays in shipment as well as any or all technical advisory/recalls/alerts, prior to or simultaneously with field alerts should be forwarded to the designated individuals determined at contract award. The Contractor will assist with a holding tank for sites that use on demand orders. Contractor will work with the laboratory to determine adequate reagent volumes for contracted period. Volumes must allow for use of the same lot number of product for the contracted period. Special handling for emergency orders of supplies: In the event that the supplies are found to be defective and unsuitable for use with the Contractor’s equipment, or the Contractor has failed to comply with the requirements for routine supply delivery, the Contractor is required to deliver the supplies within 24 hours of receipt of a verbal order for emergency delivery. If either circumstance has occurred, the Contractor shall deliver to the Government site in the most expeditious manner possible without additional cost to the Government, the necessary consumables in sufficient quantity as required to allow operation of the Contractor’s equipment for one week (under normal government test load volume). If additional requests for emergency supply delivery are required by the government, they shall be honored by the Contractor until the arrival at the laboratory of the monthly standing order/routine supplies delivery. DEFINITIONS Cost per Patient Reportable Result (CPRR) - The per patient reportable result price shall include costs covering: (1) 5 year equipment use, (2) all reagents, standards, quality controls, supplies, consumable/disposable items,...","url":"https://matchawards.com/department_of_veterans_affairs/posts/114202628878544495","identifier":"36C25525Q01820002","category":"334516 - Analytical Laboratory Instrument Manufacturing","additionalProperty":[{"@type":"PropertyValue","name":"Status","value":"Closed"},{"@type":"PropertyValue","propertyID":"PSC","name":"PSC","value":"6640 - LABORATORY EQUIPMENT AND SUPPLIES"},{"@type":"PropertyValue","propertyID":"CM category","name":"Category Management category","value":"5.4 Test \u0026 Measurement Supplies","url":"https://matchawards.com/categories/industrial-products-and-services/test-and-measurement-supplies"}],"availabilityStarts":"2025-03-21","availabilityEnds":"2025-03-28","validThrough":"2025-03-28T23:59:59-04:00","areaServed":"Kansas City, Missouri 64128, UNITED STATES","sameAs":"https://sam.gov/opp/e15e0fcf6ec14e4cbe1673f7960293ae/view"}
{"@context":"https://schema.org","@type":"BreadcrumbList","itemListElement":[{"@type":"ListItem","position":1,"name":"Home","item":"https://matchawards.com/"},{"@type":"ListItem","position":2,"name":"Contracts","item":"https://matchawards.com/contracts"},{"@type":"ListItem","position":3,"name":"Manufacturing","item":"https://matchawards.com/contracts/manufacturing"}]}
```