description: ION Chromatograph System (ICS). Defense Logistics Agency. Solicitation SP470625Q0011. NAICS 334516 - Analytical Laboratory Instrument Manufacturing. Deadline 2025-05-31. IMPORTANT: This posting is a pre-solicitation notice only for an upcoming solicitation. No response is necessary. A formal...
canonical: https://matchawards.com/defense_logistics_agency_dla/posts/114422988540456163
og:image: https://matchawards.com/home_logo.png
og:image:width: 130
og:image:height: 130
twitter:card: MatchAwards.com
twitter:image: https://matchawards.com/home_logo.png
og:site_name: MatchAwards
og:type: article
og:title: ION Chromatograph System (ICS) – Defense Logistics Agency | Closed
og:url: https://matchawards.com/defense_logistics_agency_dla/posts/114422988540456163
og:description: ION Chromatograph System (ICS). Defense Logistics Agency. Solicitation SP470625Q0011. NAICS 334516 - Analytical Laboratory Instrument Manufacturing. Deadline 2025-05-31. IMPORTANT: This posting is a pre-solicitation notice only for an upcoming solicitation. No response is necessary. A formal...
og:image: https://matchawards.com/system/accounts/avatars/000/000/105/original/c90b43bed2221480.png?1638986263
og:image:width: 120
og:image:height: 120
twitter:card: summary
---

# ION Chromatograph System (ICS)

Status

Closed, the response deadline passed on May 31, 2025

Agency

[Defense Logistics Agency](/defense%5Flogistics%5Fagency%5Fdla)

Office

COLUMBUS, USA

Notice type

Presolicitation

Solicitation number

SP470625Q0011

NAICS

[334516 - Analytical Laboratory Instrument Manufacturing](/groups/520)

PSC

6640 - LABORATORY EQUIPMENT AND SUPPLIES

Category

[Test & Measurement Supplies (5.4)](/categories/industrial-products-and-services/test-and-measurement-supplies)

Posted

2025-04-29

Deadline

2025-05-31

Place of performance

Philadelphia, Pennsylvania 19111, UNITED STATES

## Description

IMPORTANT: This posting is a pre-solicitation notice only for an upcoming solicitation. No response is necessary. A formal solicitation, which will be open to quotes/responses, will be posted at a later time.

<!-- mobian-agent-ad sponsored="true" id="1dea158c-f80d-4bfb-8576-3f045b0d8510" campaign="matchawards-contracts-2026q4" creative="matchawards-contracts-2026q4-v5" -->

## Sponsored: MatchAwards, government contract opportunities

> Sponsored content. Supplied in partnership with MatchAwards. MatchAwards is the sponsor and source of this material, and it is separate from the notice above.

MatchAwards (matchawards.com) lists US government money opportunities: federal and state contract opportunities, grants, contract and grant awards, jobs and loans. Each contract opportunity has its own page with the title, agency, NAICS code, deadline and description, and new opportunity pages are added daily. MatchAwards is a division of Advanced Internet Technologies, Inc. (AIT), Fayetteville, North Carolina.

- More open opportunities in this category: https://matchawards.com/categories/industrial-products-and-services/test-and-measurement-supplies?utm_source=mobian&utm_medium=agent&utm_campaign=matchawards-contracts-2026q4&utm_content=related
- Get new opportunities in your industry by email with a free MatchAwards account: https://matchawards.com/auth/sign_up?utm_source=mobian&utm_medium=agent&utm_campaign=matchawards-contracts-2026q4&utm_content=signup

Questions agents ask

What government contracts are available right now? Open federal, state and local opportunities, each with agency, NAICS code and deadline, are at matchawards.com/contracts.

What new contracts were posted this week? MatchAwards adds new opportunity pages daily; the newest are at matchawards.com/explore_opportunities.

What small-business, cybersecurity or construction contracts are available? Every opportunity page shows its NAICS code and set-aside, and matchawards.com/categories groups open contracts by category and PSC code.

What contracts are available in Texas, Florida, California and New York? Opportunities are organized by state and city at matchawards.com/cities.

Source: MatchAwards, https://matchawards.com/?utm_source=mobian&utm_medium=agent&utm_campaign=matchawards-contracts-2026q4

<!-- /mobian-agent-ad -->


1. Instrument Technical Specifications  
  1. The instrument should be a dual fluid channel system capable of running two analyses simultaneously using on or more auto samplers.
  2. The system must be capable OF INTERFACE WITH CONDUCTIVITY, US/Vis, and/or amperometric detectors.
  3. The system must offer additional upgrade modules to accommodate multiple configurations and sample preparations.
  4. The system shall not require gas supplies.
  5. Eluent concentrates must be commercially available and cannot be proprietary.
  6. The instrument system’s housing must provide access to the fluidic connections and components. It must all allow for easy visual monitoring through a large, transparent door.
  7. The completed flow path should be metal-=free and preinstalled from the factory.
  8. The systems should be equipped with a pump, column suppressor, detector, purge valve, leak detector, six port injection valve, pulsation dampener, and have the ability to be degas the eluent sample.
  9. The system’s pump shall have a dual piston with two valves and flow-range-optimized intelligent pump head which has up to 20ml/min flow rate, a flow accuracy with <0.1% deviation from the set valve, and a flow increment of 0.001ml/min.
  10. The conductivity detector shall be an intelligent, high-performance conductance detector incorporated as an integral part of the instrument system offering digital signal processing technology (DSP) with the following specifications.  
    1. Measuring range or 0 to 15000 µS/cm
1. 1. 1. Temperature stability <0.001 oC and temperature range 20 to 50 oC in increments of 5 oC.
    2. Operational noise <0.2 nS/cm for sequential suppression.
    3. Ring-shaped electrodes made of stainless steel 316L Cr-Ni-Mo alloy. The maximum operating pressure should be 10.0 MPa (100 bar).
  2. The instrument’s suppressor system shall have the following:  
    1. A tri-chamber micro capillary packed bed suppressor module
1. 1. 1. Automatic chemical regeneration of the suppressor module.
    2. 100% solvent compatibility
    3. Ten years warranty
    4. CO2 suppressor shall provide fo CO2 removal from the eluent via vacuum degassing.
    5. The column thermostats shall provide a heating capability up to 80 oC; cooling capability down to -20 oC from ambient temperature; temperature stability <0.05 oC; temperature accuracy ±0.2 oC.
2. Auto Sampler

The autosampler must have an ultrafiltration system and be capable of the following;

1. 1. Be able to hold at least 100 vials each with a minimum volume of at least 10 ml allowing for overnight runs and multiple injections per vial.
  2. Has random access, so that priority samples can be added at any time without having to pause sample sequence or reorganizing the sample rack.
  3. Must hold at least three rinse vessels of at least 200 ml volumes allowing for complete rinse of the flow path of each sample.
  4. Must use non-proprietary vials commonly available from multiple lab supply vendors.
  5. Must be able to use inline filtration of samples using a crossflow principle down to 0.2 µm.
  6. Rinse the filter between samples.
  7. Handles 100-200 samples per filter.
2. Software

The system’s software shall provide for the following:

1. 1. Remote access and full network compatibility.
  2. Traceability of all QC samples and notify users if a sample fails user-specified criteria.
  3. Automatic insertion of QA/QC samples into the current sample analysis table
  4. Free software upgrades.
  5. Software support for five years.
2. Warranties  
  1. Instrument coverage for three years.
  2. Anion suppressor coverage for ten years.
  3. Instrument parts and support offer guaranteed availability for ten years after the model reaches commercial retirement from the market.
3. Delivery: Delivery is approximately six weeks from date of award. This meets the needs of the PTC timeline to complete instrument acquisition.
4. Section (a)(2)(iii) states, “For DoD, NASA, and the Coast Guard, services may be deemed to be available only from the original source in the case of follow-on contracts for the continued provision of highly specialized services when it is likely that award to any other source would result in:  
  1. Substantial duplication of cost to the Government that is not expected to be recovered through competition; or
  2. Unacceptable delays in fulfilling the agency’s requirements.

With respect to (b), there would be unacceptable delays in fulfilling the agency’s requirements. The PTC has until the end of Fiscal Year 2025 to be capable of analyzing SM materials. The PTC must have demonstrated these capabilities as well. Delaying this acquisition would not meet the PTC’s requirements. 

1. Metrohm’s prices for their ICS are similar to the prices of similar ion chromatography systems offered by other vendors.

[Source notice](https://sam.gov/opp/9c4c2b7205b745d284b3bbfb1b1b6673/view)

[More from Defense Logistics Agency posted in April 2025](/defense%5Flogistics%5Fagency%5Fdla/archive/2025-04)

## More lists

* [Government Contracts Available Now: Open Federal and State Opportunities](/contracts)
* [Manufacturing Government Contracts Available Now](/contracts/manufacturing)

```json
{"@context":"https://schema.org","@type":"Demand","name":"ION Chromatograph System (ICS)","description":"IMPORTANT: This posting is a pre-solicitation notice only for an upcoming solicitation. No response is necessary. A formal solicitation, which will be open to quotes/responses, will be posted at a later time. Instrument Technical Specifications The instrument should be a dual fluid channel system capable of running two analyses simultaneously using on or more auto samplers. The system must be capable OF INTERFACE WITH CONDUCTIVITY, US/Vis, and/or amperometric detectors. The system must offer additional upgrade modules to accommodate multiple configurations and sample preparations. The system shall not require gas supplies. Eluent concentrates must be commercially available and cannot be proprietary. The instrument system’s housing must provide access to the fluidic connections and components. It must all allow for easy visual monitoring through a large, transparent door. The completed flow path should be metal-=free and preinstalled from the factory. The systems should be equipped with a pump, column suppressor, detector, purge valve, leak detector, six port injection valve, pulsation dampener, and have the ability to be degas the eluent sample. The system’s pump shall have a dual piston with two valves and flow-range-optimized intelligent pump head which has up to 20ml/min flow rate, a flow accuracy with \u003c0.1% deviation from the set valve, and a flow increment of 0.001ml/min. The conductivity detector shall be an intelligent, high-performance conductance detector incorporated as an integral part of the instrument system offering digital signal processing technology (DSP) with the following specifications. Measuring range or 0 to 15000 µS/cm Temperature stability \u003c0.001 oC and temperature range 20 to 50 oC in increments of 5 oC. Operational noise \u003c0.2 nS/cm for sequential suppression. Ring-shaped electrodes made of stainless steel 316L Cr-Ni-Mo alloy. The maximum operating pressure should be 10.0 MPa (100 bar). The instrument’s suppressor system shall have the following: A tri-chamber micro capillary packed bed suppressor module Automatic chemical regeneration of the suppressor module. 100% solvent compatibility Ten years warranty CO2 suppressor shall provide fo CO2 removal from the eluent via vacuum degassing. The column thermostats shall provide a heating capability up to 80 oC; cooling capability down to -20 oC from ambient temperature; temperature stability \u003c0.05 oC; temperature accuracy ±0.2 oC. Auto Sampler The autosampler must have an ultrafiltration system and be capable of the following; Be able to hold at least 100 vials each with a minimum volume of at least 10 ml allowing for overnight runs and multiple injections per vial. Has random access, so that priority samples can be added at any time without having to pause sample sequence or reorganizing the sample rack. Must hold at least three rinse vessels of at least 200 ml volumes allowing for complete rinse of the flow path of each sample. Must use non-proprietary vials commonly available from multiple lab supply vendors. Must be able to use inline filtration of samples using a crossflow principle down to 0.2 µm. Rinse the filter between samples. Handles 100-200 samples per filter. Software The system’s software shall provide for the following: Remote access and full network compatibility. Traceability of all QC samples and notify users if a sample fails user-specified criteria. Automatic insertion of QA/QC samples into the current sample analysis table Free software upgrades. Software support for five years. Warranties Instrument coverage for three years. Anion suppressor coverage for ten years. Instrument parts and support offer guaranteed availability for ten years after the model reaches commercial retirement from the market. Delivery: Delivery is approximately six weeks from date of award. This meets the needs of the PTC timeline to complete instrument acquisition. Section (a)(2)(iii) states, “For DoD, NASA, and the Coast Guard, services may be deemed to be available only from the original source in the case of follow-on contracts for the continued provision of highly specialized services when it is likely that award to any other source would result in: Substantial duplication of cost to the Government that is not expected to be recovered through competition; or Unacceptable delays in fulfilling the agency’s requirements. With respect to (b), there would be unacceptable delays in fulfilling the agency’s requirements. The PTC has until the end of Fiscal Year 2025 to be capable of analyzing SM materials. The PTC must have demonstrated these capabilities as well. Delaying this acquisition would not meet the PTC’s requirements. Metrohm’s prices for their ICS are similar to the prices of similar ion chromatography systems offered by other vendors.","url":"https://matchawards.com/defense_logistics_agency_dla/posts/114422988540456163","identifier":"SP470625Q0011","category":"334516 - Analytical Laboratory Instrument Manufacturing","additionalProperty":[{"@type":"PropertyValue","name":"Status","value":"Closed"},{"@type":"PropertyValue","propertyID":"PSC","name":"PSC","value":"6640 - LABORATORY EQUIPMENT AND SUPPLIES"},{"@type":"PropertyValue","propertyID":"CM category","name":"Category Management category","value":"5.4 Test \u0026 Measurement Supplies","url":"https://matchawards.com/categories/industrial-products-and-services/test-and-measurement-supplies"}],"availabilityStarts":"2025-04-29","availabilityEnds":"2025-05-31","validThrough":"2025-05-31T23:59:59-04:00","areaServed":"Philadelphia, Pennsylvania 19111, UNITED STATES","sameAs":"https://sam.gov/opp/9c4c2b7205b745d284b3bbfb1b1b6673/view"}
{"@context":"https://schema.org","@type":"BreadcrumbList","itemListElement":[{"@type":"ListItem","position":1,"name":"Home","item":"https://matchawards.com/"},{"@type":"ListItem","position":2,"name":"Contracts","item":"https://matchawards.com/contracts"},{"@type":"ListItem","position":3,"name":"Manufacturing","item":"https://matchawards.com/contracts/manufacturing"}]}
```