1. STATEMENT OF WORK

The contractor shall furnish all necessary services, labor, travel, transportation, materials, shipping, parts, connections, and equipment. This includes the provision and on-site installation of at least 5 AMSCO 50 Series Reprocessing Sinks at 4801 Linwood Blvd, Kansas City MO 64128. These reprocessing sinks are required by the Sterile Processing Services (SPS) to facilitate the manual cleaning of sterilization-compatible reusable medical devices (RMD) in invasive procedures. Their use will support a range of medical procedures and surgeries, including those in the operating room, dental services, procedural gastroenterology, procedural pulmonology and surgical clinics.

The contractor bears the responsibility for the removal and disposal of all debris and packaging materials resulting from the reprocessing sinks installation. Prior to equipment utilization, the contractor is required to provide training to clinical staff, with all duty tours for the sterile processing department being documented.

The scope of this contract comprises five AMSCO 50 Series Reprocessing Sinks. Associated services include complete installation components, execution of a post-installation check-out test, and implementation of any necessary adjustments.

2. SPECIFICATIONS

  1. Connection Type/Size- 3/8” COMPRESSION
  2. Pressure Range- 15 to 125PSI
  3. Operating Flow Rate- 1.6GPM @ MAX PSI
  4. Temperature Range 40° to 180°F
  5. Water Quality- Steris recommends the following water quality parameters:
  1. Connection Type/Size- 3/8” COMPRESSION
  2. Pressure Range- 15 to 125PSI
  3. Operating Flow Rate- 1.6GPM @ MAX PSI
  4. Temperature Range- 70° F MAX
  5. Water Quality- Steris Recommends The Following Water Quality Parameters:
  1. Connection Type/Size- ½” NPT
  2. Pressure Range- 20 TO 125 PSI
  3. Operating Flow Rate- .25 TO 2.0 GPM 15%
  4. Temperature Range- 40° F to 140° F
    • WATER GUN (WG)
  1. Connection type/size- 3/8” Compression
  2. Pressure Range- 0 TO 44 PSI
  3. Temperature Range- 40° TO 125°
  4. Water Quality- Recommendations listed under #5 of (HW & CW)
    • Air Gun (AG)
  1. Connection Type/Size- ¼” NPT
  2. Pressure Range- 30 PSI MAX
  3. Tip 1-5- Tip consumption at 30 PSI (Tip 4 and 5 must be operated at 30 PSI or less to comply with OSHA Directive.
    • Basin Drain (BD)
  1. Connection Type- 1-1/2” ABS P-TRAP (INCLUDED), 2” Copper P-Trap (Accessory)
  2. Drain Rate At 20 GPM Each Basin (Estimated)
    • Each Drain Individually Trapped
  1. When Traps Are Not Required, Elbows Can Be Purchased Locally.
  1. Weight Capacity- 562LBS (2500 N) Each
  2. Adjustment Range- 8.0” (203.2)
  3. Lift Speed- Up To 3/8” (10) Per Second
  4. Voltage- 120VAC, 3A With Overload Protection
  1. Shipping Dimensions- 132Lx 36Wx 48H
  2. Shipping Weight- 580LBS
  3. Operating Dimensions L x W x H- 97.5” x 28” x 44-9/16” TO 52-9/16” [2477 x 711 x 1132 TO 1412]

3. REQUIREMENTS

4. EQUIPMENT MAINTENANCE & RESPONSE TIMES

The vendor shall provide full-service maintenance through a service agreement for minimum 1 year post installation. The vendor shall provide all additional recurring and preventive maintenance activities:

  1. Emergency response for equipment and MSPS failures. The vendor shall respond to the facility for any equipment failure resulting in a total or partial loss of the equipment installed within or any portion of the mobile sterile processing system within 4 hours from the time of VAMC contact. If a permanent solution cannot be resolved within 24 hours of initial failure, a temporary solution shall be operational within 24 hours of initial failure. An agreed permanent solution shall be planned, and replacements ordered within 5 business days of initial failure.

5. SAFETY

The contractor/vendor shall always, ensure that all work provided complies with all local, State, and Federal rules and regulations pertaining to workplace safety. The contractor shall take all proper safety and health precautions to protect the work, the workers, the public and the property (OSHA). Installs and construction work will be in accordance with all applicable codes to include the most current codes National Fire Protection Association (NFPA), and the VA Design Guide Manuals that are applicable to the work. If the design guide is equivalent or more stringent then the mentioned codes, the VA Design Guide takes precedence.

6. DOCUMENTATION

The contractor shall furnish to the KCVA Medical Center Biomedical Engineering department legible hard copies and electronic copies of service reports of the maintenance indicating details of the maintenance performed upon completion of all work.

7. PLACE OF PERFORMANCE

Kansas City Veteran Administration Medical Center, 4801 Linwood Blvd. Kansas City 64128

8. PERFORMANCE MONITORING

The contractor’s performance will be monitored by the Contracting Officer (CO), Contracting Officer Representative (COR), Infection Control Risk Assessment Office, and Safety Office during the rental agreement period.

Coordinate all administrative and construction requirements exclusively with or through assigned Contracting Officer Representative (COR).

The Construction Safety Officer, Construction Safety Committee, and COR will review the operations and site construction conditions of the project to confirm that all Safety regulations relating to construction and healthcare facilities are being adhered to. Daily and weekly safety rounds will be conducted. For non-compliance safety related issues, a safety report will be sent to the Contractor to correct within 14 days. Non-compliance may result in job-site suspense until compliance is met at the contractor’s expense.

9. HOURS OF OPERATION

The contractor will perform services required during the hours of 7:30 am – 4:00 pm, Monday through Friday or as directed by the designated COR.